"Direct water samples for screening or an EPA approved hexane oil extraction method may be used.","The FluoroCheck II offers a fast and easy approach to the measurement of hydrocarbon oils in water. Filtered light energy targets the aromatic component of the water sample and measures the hydrocarbon fluorescence. Through an Arjay or site specific calibration, this aromatic tag is correlated to a total oil reading of your sample.","Arjay offers two test modes in one instrument; letting you determine the most suitable approach for your testing needs. You can choose from the factory preset calibrations for immediate on-site use or do your own calibration specific to your site. You can even run routine solvent-free tests and easily switch to a solvent extraction when a more selective test is warranted. All tests use disposable cuvets to eliminate cross-contamination and cleaning issues."
特性
- wastewater and groundwater
- industrial effluents
- cooling water
- produced water
技术参数
规格项 | 参数值 |
---|---|
产品类别 | Water Quality Photometers and Colorimeters |
显示 | Digital Readouts |
高量程 | 0.6464 mg/L |
仪表类型 | Portable |
本地接口 | Digital Front Panel |
低量程 | 0.0 mg/L |
测量参量 | Total Petroleum Hydrocarbons |
工作温度 | 59 to 104 F (15 to 40 C) |