Honeywell MOX-3 系列 医用氧气传感器

品牌 Sinocom 铭之光科技

分类 气体传感器

  • 产品详情
  • 附件下载

Technical SpecificationsPartial Pressure Electrochemical0-1500 mBar O29 - 13 mV in 210 mBar O2<15 s (air to 100% O2)<200 µVLinear 0-100% O2 See Note 1Product DimensionsAll dimensions in mmAll tolerances ±0.15 mm unless othewise stated Oxygen (O2) Gas SensorPart Number: AA829-M10MEASUREMENTOperating PrincipleMeasurement RangeOuput Response Time (T90)Baseline OffsetLinearity <2% O2 equivalent (0°C to 40°C) 10 kΩ Minimum (see important note)UK Data Socket AMP P/N 520257-2ELECTRICALTemperature CompensationExternal Load ResistorConnector42g (nominal)White ABSAnyMECHANICALWeightHousing MaterialOrientationCritical Care Anaesthesia-20°C to +50°C0.5 - 2.0 Bar0 - 99% RH non-condensingENVIRONMENTALTypical ApplicationsOperating Temperature Range Operating Pressure RangeOperating Humidity RangeKey Features & Benefits:• Meets the requirements of ISO 80601-2-55• Linear output from 0% to 100% O2Note 1: The regression coefficient of the best fit line should be better than 0.9995 when measured through four data points from testing with 100% N2, 21% O2, 60% O2 and 100% O2. LIFETIME< 5% signal loss/year-10°C to +40°C (short excursions to +50°C allowed)1.5 x 106 % O2hours at 20°C0.8 x 106 % O2 hours at 40°CSealed blister13 months from date of despatch(this amounts to a variation of condition 1 of our standard terms and conditions which otherwise apply)Long Term Ouput Drift in 100% O2Recommended Storage TempExpected Operating LifePackagingStandard Warranty

分享
收藏
拨打电话
立即沟通

1.点击右上角

2.分享到“朋友圈”或“发送给好友”

×

微信扫一扫,分享到朋友圈

推荐使用浏览器内置分享功能

  • #{subject}
    #{size}
    下载